Scaling up in chromatography using Linear Flow Rate

Scale-up is often an issue with protein purification. Preliminary runs are performed on small columns and then often larger amounts of product are purified on larger columns.

Because protein purification is usually performed in volume mode and not time, when scaling up it is easier to calculate the extra flow rate required for the larger column using a Linear Flow Rate, at cm/hour. This takes into account the increase in the cross-sectional area of the larger column so that it will have identical flow characteristics to the small column. Attached are 2 conversion tables for the finding of Linear Flow Rate from Volumetric Flow Rate and the other way around.

Use these two tables to help you convert the Volumetric Flow Rate to the Linear Flow Rate and vice versa.

Scaling up from Analytical to Preparative HPLC columns

When scaling up HPLC columns from analytical to preparative, multiple variables need to be taken into account. These might include changes in column length, column internal diameter, particle size increase, flow rate increase, etc.

The Knauer ScaleUp Converter is a free software tool that automatically calculates the scale-up parameters when moving an analytical-scale separation method up to a preparative one.

Go to the Knauer page (and scroll down) to download this free software tool. : https://www.knauer.net/en/knauer-scaleup-converter/p14082

Calculating the linear flow rate for CIM® monoliths

Linear flow rate calculations are commonly used with traditional chromatography resins to calculate the residence times of analytes prior to scaling up. Monolithic columns are not affected by diffusion limitations, hence residence time calculations become redundant when scaling up.

With CIMmultus™ and CIM® Tube monolith columns, the linear flow rate can be calculated with the following equation:

Where F is the flow rate in mL/min, L is the length of the monolith, and Do and Di are the outer and inner diameters of the column.

For more details on calculating linear flow rates for CIM monoliths take a look at the BIA Separations website at https://www.biaseparations.com/en/technology/calculating-linear-flow-rate-for-cimr-monoliths

Table of capillary tube diameters for calculating delay volumes

In order to make sure that chromatography peaks are collected accurately, it is important to know the delay volume of the capillary tube between the monitor and the fraction collector. This is known as the time or volume delay. This can be calculated by knowing the internal diameter of the tubing and its length.

The table below shows the internal diameters of typical capillary tubing in HPLC and FPLC. Select which is the one you are using, and then note the tube volume μl/cm that applies to it. Then multiply by the length of the tubing in cm to get the total volume of the tubing in μl. For example a capillary tube with an internal diameter of 0.5 mm, that is 60 cm long, will have a volume of 117.6 μl.

Conversion Tables: mm to inches and inch fractions to mm

Conversion Table: Pressure conversion table (MPa to bar to psi)

Conversion Table: Celsius to Fahrenheit (°C to °F)